Measuring Module PT100 - ECOLOG-PRO 4PT (EN)
1. measuring module PT100 - ECOLOG-PRO 4PT
ECOLOG-PRO 4PT, art. no. 801420
Module
Measuring module for up to 4 PT100 sensors
Circuit: 4 / 3 conductor diagram
Sensor class: A, recommended
Sensor details >>Overview sensors
ID number
The ID number is used to identify the module in elproMONITOR.
Status indicator 1; module activity
green; power supply ok
blue; system bus activity
Status indicator 2; sensor status
Condition | Description |
|---|---|
dark | No channel configured |
green | All configured sensors ok |
red | Sensor error on one or more configured sensors |
short dark | Measurement |
The display for the sensor status is updated at the recording interval!
System bus
Power supply for the connected measuring modules
Communication with the connected measuring modules
Connecting hook
For a secure connection between the modules, they are fixed at the back with the connecting hook.
1.1 Device data
Feeding
Through internal system bus
Measuring range
-200 °C..+200 °C
Measuring accuracy of the module
±0.2 °C (the sensor error and the cable error are not taken into account).
Measured value resolution
±0.1 °C
Cable fault
Cable length | Cable cross-section | Shielding | Cable fault |
|---|---|---|---|
up to 30m | 4x0.34 mm2 | No | ±0.06 °C |
30m to 50m | 4x0.34 mm2 | Yes | ±0.07 °C |
50m to 80m | 4x0.34 mm2 | Yes | ±0.2 °C |
Recording interval
1 minute..60 minutes
Each channel can record with a separate interval
Date and time
Internal system clock
The internal system clock is permanently synchronized with the elproMONITOR system time