ECOLOG-PRO xG - Operation Manual

Information

Important information and warnings

DELETE

Windows function, e.g. Delete

 

<xxxxxxxx>

Syntax for placeholders.
Placeholders are written between < > characters.

 

>>

Reference to a further chapter or document.

 

The representation refers only to a part of the user interface.

 

In the interest of our clientele, we reserve the right to make changes due to technical advances. Therefore, schematics, descriptions and scope of delivery may be changed without prior notice. This manual is valid as of version 1.0

Liability

  1. In no event shall ELPRO-BUCHS AG be liable for any direct, indirect, special, incidental, consequential or punitive damages or losses, including lost profits and data, arising out of the use of your data loggers, sensors, accessories, software products or the information contained in the documentation.

  2. ELPRO-BUCHS AG does not bear any responsibility for the installation of the software products.

  3. ELPRO-BUCHS AG makes no warranties of any kind, express or implied, regarding the merchantability and fitness of your products for any particular purpose.

  4. In the interest of our clientele, we reserve the right to make changes and discontinuations due to technical advances. Therefore functionality, schemes, descriptions and scope of delivery can be changed without prior notice!

  5. Because some countries or states do not allow the limitation of the term implied warranty or disclaimer of incidental or consequential damages, the limitations and exclusions may not apply to every purchaser. If any provision of this warranty is held invalid or unenforceable by a court of competent jurisdiction, it will not affect the validity or enforceability of the remaining provisions.

  6. ELPRO-BUCHS AG accepts no liability for transport damage or any consequential damage resulting therefrom.

  7. In general, the legal provisions of Switzerland apply. The place of jurisdiction is the district court of Werdenberg-Sarganserland

Warranty

  • The warranty obligations are limited, at ELPRO-BUCHS AG's discretion, to the repair, refund of the purchase price, repair free of charge or replacement of a defective product returned to ELPRO-BUCHS AG or a reseller approved by ELPRO-BUCHS AG within the warranty period.

  • There is no entitlement to free "on-site customer support" by an employee of ELPRO-BUCHS AG.

  • The warranty extends only to the original purchaser or end user and customers of a reseller authorized by ELPRO-BUCHS AG.

  • Resellers approved by ELPRO-BUCHS AG are not authorized to provide a greater or other warranty on behalf of ELPROBUCHS AG.

  • ELPRO-BUCHS AG grants a warranty period of 24 months on the following new products:

    • Data logger

    • Brackets

    • Accessories, except sensors and third-party products

  • This warranty applies to material defects or production defects.

  • ELPRO-BUCHS AG grants a warranty period of 6 months for the following products and services:

    • All service work and repairs

    • Temperature sensors

    • Humidity sensors

  • ELPRO-BUCHS AG grants a warranty period of 90 days that the medium on which the software product was delivered is free from defects in material and workmanship under normal conditions. The software product conforms in all material respects to the operating instructions and the information in the help file.

  • The warranty does not apply to consumables, disposable batteries or any other product which, in the opinion of ELPRO-BUCHS Ltd:

    • abused

    • changed

    • replaced by a wrong product

    • has been operated and damaged outside the specification due to accident or operating or handling conditions.

  • Wear and tear, cable breakage and corrosion cannot be claimed under warranty.

  • For third-party products, ELPRO-BUCHS AG grants at most the warranty period of the manufacturer.

  • For discontinued products, ELPRO-BUCHS AG provides warranty work and customer support only for a limited period of time.

  • Warranty repairs are generally only carried out at the factory (ELPRO-BUCHS AG) or at an agency recognized by the factory.

  • ELPRO-BUCHS AG does not provide any warranty regarding the standard or SCS calibration of the data loggers and sensors. The specified data correspond to the situation during the calibration process

Software

  1. The software products of ELPRO-BUCHS AG are subject to internal quality regulations and are regularly validated within the factory. In the event of any program errors, bypassing the error is considered as elimination.

  2. Software operating instructions do not contain instructions on the basic operation of a computer or on basic functions of the Windows® operating system. For information on how to operate your computer or operating system, please refer to the appropriate manuals for your computer

Data loggers, sensors and accessories

  • ELPRO-BUCHS AG uses the highest quality standards and a certified quality management system according to ISO 9001 for the production of data loggers and their accessories.

  • For information on the operation of the data loggers and their accessories, please refer to the corresponding product documentation provided by ELPRO-BUCHS AG.

  • When installing data loggers, sensors and accessories, the locally applicable installation regulations must be observed.

  • For applications in potentially explosive atmospheres, the zone assignment specified by ELPRO-BUCHS AG as well as the application and safety instructions must be strictly observed.

  • In the event of a warranty claim, the customer will receive a repair cost estimate from ELPRO-BUCHS AG in order to issue the appropriate authorization before work commences.

  • Transport costs for a repair at ELPRO-BUCHS AG will be borne by the customer. DAP (value added tax) will be borne by ELPRO-BUCHS AG.

  • ELPRO-BUCHS AG reserves the right to charge the purchaser for costs of repairs/parts replacement.

  • After repair, the product will be returned to the buyer with return shipping charges billed to the buyer (FOB shipping location).

Trademark

  • All listed company and product names and their trademarks are the protected property of their respective owners.

  • This product is subject to CE marking.

  • The manufacturer guarantees the conformity of this product to the corresponding directives: EN 61000-6-2 : 2001 and EN 61000-6-4 : 2001

  • FCC ID: Z45-E11645398

  • IC: 9954A-E11645398

image-20240110-102226.png

 

  • This product is subject to UKCA marking.

  • The manufacturer guarantees the conformity of this product to the corresponding UKCA guidelines.

  • This product complies with the regulatory requirements specified by the Ministry of Internal Affairs and Communications (MIC) in Japan, as outlined in the applicable standards and regulations.

 

  • This product complies with the regulatory requirements specified by the Australian Communications and Media Authority (ACMA) for Radio-Networking Equipment (R-NZ), as outlined in the applicable standards and regulations.

 

  • This product complies with the regulatory requirements specified by the Infocomm Media Development Authority (IMDA) in Singapore, as outlined in the applicable standards and regulations.

  • This product must be disposed of in accordance with WEEE (Waste electrical and electronic equipment, 2002/ 96/EC)

1. Safety Instructions

Intended Use

All electrical devices produced by ELPRO are intended for commercial use („business to business“).

Environmental Conditions

Temperature

Temperatures outside operation range may damage battery. Operation ranges:

ECOLOG-PRO 1TGi, 1THGi, 1TGe: -35°C … +55°C

Water/Humidity

Limited protection against dust ingress and protected against splash water from any direction.

Pressure

Overpressure or vacuum can damage the device. Do not vacuum if used for airfreight.

Mechanical Force

Avoid violent knocks and blows.

IR radiation

Avoid exposure to IR radiation (heat and superheated steam can result in deformation of the case).

Microwave

Do not expose to microwave radiation (risk of battery explosion).

X-ray

Avoid long-term exposure to X-ray (risk of harm to device).
Tests of short x-ray exposures as part of transportation processes (airports, customs) have been performed and documented (available at ELPRO).

 

Battery

Batteries can be changed. Material safety data sheet according to provisions of directive 91/155/EEC and shipping information are available from ELPRO. Do not subject the batteries to mechanical stress nor dismantle them. The leaking battery fluid is highly corrosive and can generate severe heat when it comes into contact with moisture or it can ignite fire.

Safe Use

Ordinary people may install and operate the device without further safeguards.

Radio Equipment

This equipment emits radiated power: LTE Bands 1, 2, 3, 4, 5, 8, 12, 13, 14, 17, 18, 19, 20, 25, 26, 28, 66 maximum power: 23 dBm. Note: For Taiwan the equipment emits radiated power: LTE Bands 3, 8, 28.

2. Quick Start Guide

 

3. System Overview

The ECOLOG-PRO xG autonomous Mobile-IoT sensors with replaceable battery, communicate via public telecom network. The data logger family described in this document are used for stationary temperature and humidity monitoring. The measured values are transmitted via the cellular network to a monitoring software (elproCLOUD) which stores and analyzes the data, provides alerts if alarm limits are violated, and generates reports. The system provides superior visibility and transparency in meeting GxP requirements. The sensor-based monitoring software is easily accessible via a web browser and is also used to configure the devices.

Following pages cover key functionality of the wireless data logger (ECOLOG-PRO 1TGi / 1THGi / 1TGe). For more detailed software instructions, please visit our online knowledge base:
https://www.elpro.com/en/support/elpro-cloud

3.1 Device overview

Brand

ELPRO-BUCHS AG
Langaeulistrasse 45
9470 Buchs SG
Switzerland

Model

ECOLOG-PRO xG

ECOLOG-PRO xG Types 

 

ECOLOG-PRO 1TGi

 

ECOLOG-PRO 1THGi

 

ECOLOG-PRO 1TGe

Wireless realtime data logger, cellular data transmission (NB-IoT / LTE-M).

Data logger has a display and internal memory so that all data is available.

Stationary Use

Application range: 0 °C..+55 °C for Alkaline battery / -35 °C..+55 °C for LiFe battery
Measurement range of internal sensor: -40 °C..+70 °C

If the customer wants to use ECOLOG PRO xG in transport, it is required to replace the device annually.

internal temperature sensor

internal temperature and humidity sensor

internal and external temperature sensor

Alkaline or Lithium iron sulfide AA batteries (replaceable)

Voltage: 2 x 1.5 Volts

Alkaline or Lithium iron sulfide AA batteries (replaceable)

Voltage: 2 x 1.5 Volts

Alkaline or Lithium iron sulfide AA batteries (replaceable)

Voltage: 2 x 1.5 Volts

For detailed technical specifications: please go to http://www.elpro.com/

 

4. Functionality and Modes

Unless otherwise noted the following information applies equally to all three ECOLOG-PRO xG models. After configuration of the data logger measured values for temperature and relative humidity (ECOLOG-PRO 1THGi only) are recorded, stored and evaluated with regard to the defined alarm criteria. The display shows the current mode.

4.1 Elements

4.1.1 Generic elements

1

Display

2

Pairing button

Press long (> 2 seconds) initiates communication with cloud

Note: after input of batteries the device goes into pairing mode

3

Light sensor

if light sensor is configured, make sure it is not dirty or covered

4

QR-Code containing Device ID and weblink to cloud

5

Info / Menu button

Press short (< 1 second) = Info (toggle display / menu)

Press long (> 2 seconds) = Menu (open menu / select menu entry)

6

Device type

7

Device ID

4.1.2 Specific elements

Logger type

Element

 

ECOLOG-PRO 1TGe

M8 connector for Pt100 probe

Compare chapter 6.2 - External Pt100 Probes for ECOLOG-PRO 1TGe for available probes

 

ECOLOG-PRO 1THGi

Slot for combined temperature / humidity sensor

 

 

4.2 Display

 

Icon

Name

Description

1

No connection to Cloud

No connection to Cloud possible or communication failure

2

Run

Device is measuring and buffering

3

Signal strength

Not visible if radio off / flight-mode

4

Warning

Warning (configurable) for

  • Temperature warning limit

  • Communication warning

  • Light warning

  • Low battery warning

5

Alarm On/Off

Shows if Alarm criteria are active or paused

6

Alarm Status

Display (configurable in future versions)

OK or Alarm (pushed from Cloud)

After excursion, Alarm will rmain on display

7

Battery level

4 battery levels

8

8 Digit display

 

Various functions, e.g.

  • temperature

  • status

  • customer field (e.g. specific room or equipment name, ID, …)

4.3 States

ECOLOG-PRO xG devices are mainly used to monitor temperature sensitive products through stationary use. The device has various configuration options available. The device states are visualized below and further described in subsequent chapters. The workflow options may vary depending on configuration and device type.

4.3.1 Shelflife

When delivered, the device is not powered, but with insulation strips on the battery contacts.

  • After removing the isolation strip / inserting the batteries, the instrument asks if the batteries are new (NEW.BAT: Y/N). Confirm by pressing the Info/Menu button for more than 2 seconds.

  • Then, the instrument goes directly to CONFIG state. Through the RADIO.OFF menu the device can enter SHELFLIFE state.

 

4.3.2 Configuration & Start

  • By pressing the pairing button (>2 seconds) the device is pulling a configuration of the cloud, and the Run icon on the display indicates a successful start.

4.3.3 Pause (Alarming off)

Via the elproCLOUD software, the alarming of the device can be turned off and on again in order to pause alarming for a specific timeframe.

4.4 Menu

The ECOLOG-PRO xG family feature a menu to operate the device:

  • To enter the menu, press the Info/Menu button for at least two seconds

  • To switch between the menu options, press the Info/Menu button shortly

  • To select a menu item, press the Info/Menu button for at least two seconds. To confirm, a selected menu item blinks once.

  • To leave the menu

    • wait for 5 seconds

    • or press the left button

    • or select the last menu item EXIT

 

All menu items and their availability are described in the table below (no menu available in SHELFLIFE mode)

State


Menu item

CONFIG

TRANSIT

PAUSED

GET.CONFI

trigger communication to pull (new) configuration from cloud

 

 

 

PUSH.DATA

trigger upload of data to cloud

 

 

 

RADIO.OFF

The communication turns off

  • state changes from CONFIG to SHELFLIFE
    (configuration remains if device started again)

 

 

 

SENSORS

show current measurement for temperature, humidity, light

ABOUT

show device info (firmware version, config ID, IMEI, last communication time, battery voltage)

MARK

set marker

 

 

FCT.RESET

reset device (to be confirmed by pressing Info button again)

 

 

 

5. Further instructions

5.1 Pairing procedure

 

6. Accessories

6.1 Bracket

ELPRO offers an optional bracket (BRA_LIBERO Gx, part number 802375) for mounting of ECOLOG-PRO xG data loggers if required, i.e. to containers for cryogenic applications. The bracket is made of solid ABS plastic to protect the data logger but not influence the communication. It consists of an upper and a lower part. The ECOLOG-PRO xG is inserted into the lower holder from above.

1

Various mounting options

  • 360° screwing

  • Adhesive tape

  • Cable strap

2

Cable wire mouth

3

Transparent cover allows to read display

4

A button slot to operate the device

5

The snapping mechanism to close the cover

6

Secure fixation for the ECOLOG-PRO xG

7

Possibility to lock the cover

 

6.2 External Pt100 Probes for ECOLOG-PRO 1TGe

ECOLOG-PRO 1TGe can be used for different applications, depending on the sensor element. ELPRO offers standard probes for three main applications:

  • Cryogenic storage

  • Dry ice storage

  • Freezer (-25 °C ... -15 °C, typical) / fridge (+2 °C ... +8 °C) / ambient (+15 °C … +25 °C) storage

6.2.1 Cryogenic storage

For cryogenic applications the ECOLOG-PRO 1TGe is usually mounted directly to the container or the container lid, using the optionally available bracket with the sensor leading into the tank. ELPRO offers an easy, turnkey service for mounting the assembly and calibration.

ELPRO offers three Pt100 standard probes for cryogenic applications with M8 connector in different lengths:

 

Designation

Article Number

PRO_PT100_ST300D3_SIL0.05_M8

802287

Designation

Article Number

PRO_PT100_ST300D3_SIL0.05_M8

802287

Picture

Description

Pt100 4-wire probe, M8 connector to connect to LIBERO CE/GE, bendable once at room temperature (do not kink). typical application = CRYO

Probe material

Stainless steel

Probe length

30 cm

Probe diameter

3 mm

Temperature range of probe

-200 °C … +200 °C

-Temperature range Class A

n.a.

-Temperature range Class B

-50 °C … +200 °C

Cable material

Silicone (black)

Cable length

0.05 m

Cable diameter

4 mm

Temperature range of cable

bendable in the range between -60 °C … +90 °C

Drawing

 

 

Designation

Article Number

PRO_PT100_ST350D3_SIL_0.05_M8

802288

Designation

Article Number

PRO_PT100_ST350D3_SIL_0.05_M8

802288

Picture

Description

Pt100 4-wire probe, M8 connector to for LIBERO CE/GE/GF ext, bendable once at room temperature (do not kink). typical application = CRYO

Probe material

Stainless steel

Probe length

35 cm

Probe diameter

3 mm

Temperature range of probe

-200 °C … +200 °C

-Temperature range Class A

n.a.

-Temperature range Class B

-50 °C … +200°C

Cable material

Silicone (black)

Cable length

0.05 m

Cable diameter

4 mm

Temperature range of cable

bendable in the range between -60 °C … +90 °C

Drawing

 

 

 

Designation

Article Number

PRO_PT100_ST50D3.2_PTFE1_M8

802340

Designation

Article Number

PRO_PT100_ST50D3.2_PTFE1_M8

802340

Picture

Description

Pt100 4-wire probe, with M8 connector for LIBERO CE/GE/GF ext, waterproof, typical application = CRYO

Probe material

Stainless steel

Probe length

5 cm

Probe diameter

3.2 mm

Temperature range of probe

-200 °C … +200 °C

-Temperature range Class A

-200 °C … +200 °C

-Temperature range Class B

n.a.

Cable material

PTFE (white)

Cable length

1 m

Cable diameter

3 mm

Temperature range of cable

-200 °C … +200 °C

Drawing

 

6.2.2 Dry ice storage

Also in dry ice applications, the ECOLOG-PRO 1TGe is usually attached to the outside of the container using the optionally available bracket and the sensor leads into the tank. ELPRO offers an easy, turnkey service for mounting the assembly and calibration.

For this application, ELPRO offers two standard probes with a probe length of 10 cm and Teflon cable in different lengths:

 

 

Designation

Article Number

PRO_PT100_ST100D4_PTFE1_M8

802284

Designation

Article Number

PRO_PT100_ST100D4_PTFE1_M8

802284

Description

Pt100 4-wire probe, with M8 connector to connect to LIBERO CE/GE/GF ext, typical application = Dry Ice

Probe material

Stainless steel

Probe length

10 cm

Probe diameter

4 mm

Temperature range of probe

-90 °C ... +250 °C

-Temperature range Class A

-30 °C … +200 °C

-Temperature range Class B

-50 °C … +200 °C

Cable material

PTFE (white)

Cable length

1 m

Cable diameter

3.5 mm

Temperature range of cable

bendable in the range between -90 °C … +70 °C

Drawing

 

Designation

Article Number

PRO_PT100_ST100D4_PTFE2.65_M8

802285

Designation

Article Number

PRO_PT100_ST100D4_PTFE2.65_M8

802285

Description

Pt100 4-wire probe, with M8 connector to connect to LIBERO CE/GE/GF ext, typical application = Dry Ice

Probe material

Stainless steel

Probe length

10 cm

Probe diameter

4 mm

Temperature range of probe

-90 °C ... +250 °C

-Temperature range Class A

-30 °C … +200 °C

-Temperature range Class B

-50 °C … +200 °C

Cable material

PTFE (white)

Cable length

2.65 m

Cable diameter

3.5 mm

Temperature range of cable

bendable in the range between -90 °C … +70 °C

Drawing

 

6.2.3 Freezer, fridge, ambient storage

For temperature monitoring of freezers, refrigerators or rooms, ELPRO offers two waterproof silicon Pt100 probes with different cable lengths as standard articles:

Designation

Article Number

PRO_PT100_P20D5_PLA1_M8

802290

Designation

Article Number

PRO_PT100_P20D5_PLA1_M8

802290

Description

Pt100 4-wire probe, with M8 connector to connect to LIBERO CE/GE/GF ext, waterproof, typical application = Freezer, fridge or ambient shipments and storage

Probe material

Silicone

Probe length

2 cm

Probe diameter

5 mm

Temperature range of probe

-50 °C ... +105 °C

-Temperature range Class A

-30 °C … +105 °C

-Temperature range Class B

-50 °C … +105 °C

Cable material

Silicone (black)

Cable length

1 m

Cable diameter

4 mm

Temperature range of cable

bendable in the range between -60 °C … +90 °C

Drawing

 

Designation

Article Number

PRO_PT100_P20D5_PLA2.65_M8

802291

Designation

Article Number

PRO_PT100_P20D5_PLA2.65_M8

802291

Description

Pt100 4-wire probe, with M8 connector to connect to LIBERO CE/GE/GF ext, waterproof, typical application = Freezer, fridge or ambient shipments and storage

Probe material

Silicone

Probe length

2 cm

Probe diameter

5 mm

Temperature range of probe

-50 °C ... +105 °C

-Temperature range Class A

-30 °C … +105 °C

-Temperature range Class B

-50 °C … +105 °C

Cable material

Silicone (black)

Cable length

2.65 m

Cable diameter

4 mm

Temperature range of cable

bendable in the range between -60 °C … +90 °C

Drawing

 

6.3 Extension of sensor cables

An extension cable with two M8 connectors at a length of 1 m is also available to attach the data logger and the probe.

 ATTENTION:

Total cable length (including sensor and cable tail on the data logger) must not exceed 3 m!

 

Designation

Article Number

ECA_PLA_1M_M8

802282

Designation

Article Number

ECA_PLA_1M_M8

802282

Description

Extension cable 1 m for LIBERO CE / LIBERO GE / LIBERO GF ext. M8 connector at both ends (male / female),
PVC, -60 °C ...+90 °C.

Cable material

PVC (black)

Cable length

1 m

Cable diameter

3.5 mm

Temperature range of cable

bendable in the range between -60 °C … +90 °C

 

6.3 M8 Connector incl. mounting service

ELPRO offers a mounting service, adding an M8 connector to a Pt100 temperature sensor in order to use any 4-wire Pt100 probe in combination with ECOLOG-PRO 1TGe.

 

Designation

Article Number

CTR_M8_SER

802289

Designation

Article Number

CTR_M8_SER

802289

Drawing

 

 

7. ERROR codes & actions

Message

Description

Recommended actions

Message

Description

Recommended actions

N.C.

No connected probe. Logger cannot receive temperature values.

Connect a probe to the M8 connector.

Config Error

The configuration has not arrived at the device.

Make sure that the logger has enough signal (minimum 1 bar) and try again.

NO CONF

Device has no valid configuration

Make sure the correct ID is entered in the elproCLOUD. Make sure that the logger has enough signal (minimum 1 bar) and PULL.CONFIG again.

REJECTED

configuration could not be received (wrong ID, pairing request expired)

Make sure the correct ID is entered in the elproCLOUD. Make sure that the logger has enough signal (minimum 1 bar) and PULL.CONFIG again.

FATAL ERROR

 

Create an ELPRO Support Ticket Support & Knowledge Base | Create Support Ticket (elpro.com)

ERROR 35

 

Create an ELPRO Support Ticket Support & Knowledge Base | Create Support Ticket (elpro.com)

For further information and FAQs visit our ELPRO Support Website Support & Knowledge Base (elpro.com).

 

 

8. Disposal

a)      Device

Electronic devices are recyclable and do not belong in the household waste. Dispose of the product at the end of its service life in accordance with applicable laws. Remove any batteries and dispose of them separately from the product.

 b)      Batteries

You are legally obliged to dispose of all used batteries according to applicable laws; disposal via household waste is prohibited. Batteries are marked with the adjacent symbol, under which is printed the chemical symbol for the heavy metal (Cd = cadmium, Hg = mercury, Pb = lead). This indicates the battery contains hazardous material. You can dispose of used batteries at collection points in your local community. Please help protect our environment and dispose of batteries properly.

 

9. Battery replacement

To change the batteries, the battery cover on the back must be opened.

The batteries must not be removed axially, but must be carefully removed radially with a screwdriver (see graphic below).

Confirm that you have inserted new batteries by pressing the Info/Menu key for 3 seconds (NEW.BAT: Y).

ELPRO recommends the use of AA Ultimate Lithium L91 batteries from the manufacturer Energizer.

To order a pack of 10 pieces, please contact your local ELPRO office.

10. Declaration of Conformity

 10.1 EU Declaration

 

10.2 FCC/ISED Regulatory notices

 

10.3 NCC Regulatory notices

In order to reduce the effects of electromagnetic waves, please use the appropriate MPE standard value: 0.9 mW/cm², measured value: 0.12239 mW/cm², recommended for use with the antenna at least 20 cm from the human body.

減少電磁波影響,請妥適使用

電波功率密度 MPE 標準值:0.9 mW/cm²,送測產品實測值:0.012239 mW/cm²,建議使用時設備天線至少距離人體20公分

 

10.4 UKCA Conformity

 

 

 

ELPRO-BUCHS AG

Langäulistrasse 45
9470 Buchs
SWITZERLAND

E-Mail: swiss@elpro.com

For local agencies see:

http://www.elpro.com

 