Wireless Measuring Module ECOLOG-PRO 1NTR (EN)
1. wireless measuring module ECOLOG-PRO 1NTR
ECOLOG-PRO 1NTR 868 MHz, art. no. 801429 / ECOLOG-PRO 1NTR 915 MHz, art. no. 802251 / ECOLOG-PRO 1NTR 922 MHz, art. no. 802558
Module
ECOLOG-PRO 1NTR is a measuring module with an internal temperature sensor.
This module communicates with the ECOLOG-PRO RBR via a radio signal.
ID number
"S" followed by a 6-digit number; the ID number is used to identify the module in the monitoring software
Link; left and center LEDs
Flashing red (left); ECOLOG-PRO 1NTR is active, but not connected to a communication module (ECOLOG-PRO RBR)
Flashing green (center); ECOLOG-PRO 1NTR is active and connected to a communication module (ECOLOG-PRO RBR)
The interval of the flashing light is 10s.
Data; right LED
Flashing green (center); The measurement on the channel has started. Everything is in order.
The interval of the flashing light is 10s.
Radio connection to the communication module
Available in three versions:
868 MHz (Europe, Singapore)
915 MHz (USA, Canada, Australia, New Zealand, Argentina, Brazil)
922 MHz (South Korea)
1.1 Device data
Feeding
2 x AAA alkaline batteries or
2 x LiFe batteries (lithium-iron)
Improper use or misuse of batteries can lead to leakage of the battery fluid or, in the
In the worst case, this could lead to an explosion and/or fire.Remove the batteries if you will not be using the device for a long period of time to avoid damage due to leakage. Leaking or damaged batteries can cause acid burns on contact with the skin. Wear protective gloves when handling damaged batteries.
Remove expired batteries from the device immediately.
Do not take the batteries apart, do not short-circuit them and do not throw them into a fire. Do not throw them into a fire. Never attempt to charge non-rechargeable batteries. There is a risk of explosion!
Keep batteries out of the reach of children. Do not leave batteries lying around as they can be swallowed by children or pets.
Always select the correct battery type that is best suited to your intended use (see 'Measuring range').
To ensure that the battery status can be determined correctly, always insert new, unused batteries when changing the batteries.
Never insert different types of batteries into the device at the same time.
All batteries should be replaced at the same time. Mixing old and new batteries can lead to battery leakage and damage the device.
Make sure that the battery contacts and the mating contacts in the device are clean before inserting the batteries.
Ensure that the batteries are inserted correctly, i.e. observe the polarity (+ and -).
Battery life
14 months (provided that the radio connection of the module is stable and the defined measuring range is maintained)
Measuring range
With alkaline batteries: -10 °C.. +50 °C
With LiFe batteries: -35 °C..+50 °C
Measuring accuracy of the module
±1.0 °C in the range between -35.0 °C and -10.1 °C
±0.5 °C in the range between -10.0 °C and -0.1 °C
±0.4 °C in the range between 0.0 °C and +25.0 °C
±0.8 °C in the range between +25.1 °C and +50.0 °C
Measured value resolution
0.1 °C
Recording interval
1 - 60 minutes; user-programmable via the monitoring software
The recording interval has an impact on battery life
Channels
1 temperature channel
Sensor type