Measuring Module T/rH - ECOLOG-PRO 2TH (EN)
1st measuring module T/rH - ECOLOG-PRO 2TH
ECOLOG-PRO 4PT, art. no. 801424
Module
Measuring module for up to 2 T/rh sensors
ID number
The ID number is used to identify the module in elproMONITOR.
Status indicator 1; module activity
green; power supply ok
blue; system bus activity
Status indicator 2; sensor status
Condition | Description |
---|---|
dark | No channel configured |
green | All configured sensors ok |
red | Sensor error on one or more configured sensors |
short dark | Measurement |
System bus
Power supply for the connected measuring modules
Communication with the connected measuring modules
Connecting hook
For a secure connection between the modules, they are fixed at the back with the connecting hook.
1.1 Device data
Feeding
Through internal system bus
Measuring range
Depending on the application, you can choose between 2 sensor variants:
Temperature / humidity sensor, electrolytic (-20 °C..+80 °C, 0 %rH..100%rH) >> 802363 - PRO_TRH_nSens-HT-EIS - Temperatur / Feuchte Sensor, elektrolytisch
Environments with aggressive gases at high temperature and humidity (stability chamber, incubators)
Low temperature environment (refrigerator)
Measurements with a high accuracy of ±0.5 %rH
Temperature / humidity sensor, capacitive (-20 °C..+80 °C, 20 %rH..80%rH) >> 802209 - PRO_TRH_nSens-HT-CSP - Temperatur / Feuchte Sensor, kapazitiv
This sensor is suitable for monitoring storage rooms, laboratories or general room climate monitoring
Connection cable lengths
5m
10m
30m
Recording interval
1 minute..60 minutes
The two measured values temperature and humidity are recorded synchronously for each sensor. The interval for the two sensors together can be selected individually.
Date and time
Internal system clock
The internal system clock is permanently synchronized with the elproMONITOR system time
Data security
Internal, non-volatile memory
If the network connection fails, the measured values are still recorded and temporarily stored in the internal memory
10'000 measured values
Scope of delivery
Measuring module
Connecting hooks, see back of module
Info sheet
Validation certificate
The plug connectors for the sensor connection are not included in the scope of delivery.
1.2 Wiring diagram
The sensor connection is made via a 3-pin standardized plug connection, which is implemented on both the extension cables and the sensors.
Channel | Sensor | Measured value - connection type |
---|---|---|
CH1 | Sensor 1 | Temperature - electrolytic or capacitive sensor |
CH2 | Sensor 1 | Humidity - electrolytic or capacitive sensor |
CH3 | Sensor 2 | Temperature - electrolytic or capacitive sensor |
CH4 | Sensor 2 | Humidity - electrolytic or capacitive sensor |
Clamp | Occupancy |
---|---|
1 | Feeding |
2 | Signal |
3 | Mass |