Wireless Measuring Module T/rF ECOLOG-PRO 1THR (EN)
1. wireless measuring module T/rF ECOLOG-PRO 1THR
ECOLOG-PRO 1THR 868 MHz, art. no. 802268 / ECOLOG-PRO 1THR 915 MHz, art. no. 802269 / ECOLOG-PRO 1THR 922 MHz, art. no. 802559
Module
ECOLOG-PRO 1THR is a measuring module with an external combination sensor for temperature and relative humidity.
This module communicates with the ECOLOG-PRO RBR via a radio signal.
ID number
"R" followed by a 6-digit number; the ID number is used to identify the module in the monitoring software
Link; left LEDs
Flashing red; ECOLOG-PRO 1THR is active, but not connected to a communication module (ECOLOG-PRO RBR)
Flashing green (10s); ECOLOG-PRO 1THR is active and connected to a communication module (ECOLOG-PRO RBR)
yellow flashing (1s); ECOLOG-PRO 1THR is currently receiving its registration information. Please wait for the green light.
Flashing yellow (10s); ECOLOG-PRO 1THR is active, but registration was not successful and must be restarted by briefly pressing the pairing button Button (less than 1s)
Data; right LED
Red flashing light; measurement has started on one or both channels. One or more errors have occurred during the measurement.
Flashing green (10s); The measurement on the channel has started. Everything is OK.
Radio connection to the communication module
Available in three versions:
868 MHz (Europe, Singapore)
915 MHz (USA, Canada, Australia, New Zealand, Argentina, Brazil)
922 MHz (South Korea)
1.1 Device data
Feeding
3 x AA alkaline batteries or
Improper use or misuse of batteries can lead to leakage of the battery fluid or, in the
In the worst case, this could lead to an explosion and/or fire.Remove the batteries if you will not be using the device for a long period of time to avoid damage due to leakage. Leaking or damaged batteries can cause acid burns on contact with the skin. Wear protective gloves when handling damaged batteries.
Remove expired batteries from the device immediately.
Do not take the batteries apart, do not short-circuit them and do not throw them into a fire. Do not throw them into a fire. Never attempt to charge non-rechargeable batteries. There is a risk of explosion!
Keep batteries out of the reach of children. Do not leave batteries lying around as they can be swallowed by children or pets.
Always select the correct battery type that is best suited to your intended use (see 'Measuring range').
To ensure that the battery status can be determined correctly, always insert new, unused batteries when changing the batteries.
Never insert different types of batteries into the device at the same time.
All batteries should be replaced at the same time. Mixing old and new batteries can lead to battery leakage and damage the device.
Make sure that the battery contacts and the mating contacts in the device are clean before inserting the batteries.
Ensure that the batteries are inserted correctly, i.e. observe the polarity (+ and -).
Battery life
14 months (provided that the radio connection of the module is stable and the defined measuring range is maintained)
Measuring range
Depending on the application, you can choose between 2 sensor variants:
Temperature / humidity sensor, electrolytic (-20 °C..+80 °C, 0 %rH..100%rH) >> 802363 - PRO_TRH_nSens-HT-EIS - Temperature / humidity sensor, electrolytic
Environments with aggressive gases at high temperature and humidity (stability chamber, incubators)
Low temperature environment (refrigerator)
Measurements with a high accuracy of ±0.5 %rH
Temperature / humidity sensor, capacitive (-20 °C..+80 °C, 20 %rH..80%rH) >> 802209 - PRO_TRH_nSens-HT-CSS - Temperature / humidity sensor, capacitive
This sensor is suitable for monitoring storage rooms, laboratories or general room climate monitoring
Measuring accuracy of the module
±0,2 °C
Measured value resolution
0.1 °C
Cable length
5m
10m
Recording interval
1 - 60 minutes; user-programmable via the monitoring software
The two measured values temperature and humidity are recorded synchronously for each sensor. The interval for the two sensors together can be selected individually.
The recording interval has an effect on the battery life